Contact an Account ManagerAssay of paclitaxel (taxol) in plasma and urine
Summary
Type: HPLC
Application #: 110310
Catalog#: 720014.46 - MN HPLC Column, Nucleosil C18, 100A, 5um, 4.6x250mm
Detection: UV, 227 nm
Conditions: Eluents: Acetonitrile / 35 mmol ammonium acetate buffer / THF 45:50:5 (v/v/v)
Flow rate: 1.8 mL/min
Authors: Martin N
Source: J of Chromatography B 709 (1998) 281-288